The Keystone system is the result of Pulmonary Data Services 20+ years of experience in building and advancing the technology of pulmonary function analyzers. This system is a simple, inexpensive but comprehensive analyzer with advanced software for total automation and control of lung function measurements. The keystone uses a direct displacement spirometers. Lung volume measurements are determined by the steady state helium method and diffusion analysis is determined by the single-breath method.
Features
- Totally automated test sequencing
- Single pneumatic valve for all procedures
- Simultaneous, direct pen-linked and computerized graphing
- 100% sterilization ability for the breathing circuit and valve
- Menu driven software with on-line help
- Automated leak detection, calibration and linearity verification
- Self-calibrating gas analyzer
- on-screen display of QC data during test procedures
- User definable:
- Report Formats
- predicted values
- screen and report graphics
- automated test procedures
- trend analysis
- Multiple graphic overlays
- Exceeds all published standards – ATS, NIOSH, OSHA and Social Security
Specifications
Spirometer type dry rolling seal
Measuring Range 0-10 Liters
Accuracy ± or 50ml over 10 liter range
Resistance < H2O mm/sec direct pen-linkage strip chart recorder
Helium Analyzer Type Thermal conductivity
Carbon Monolide Analyzer Type Electrochemical cell
Electical Voltage Requirements 120 vac/5 amp
Leakage Current less than 100 micro-amperes A/D Converter
Dimensions 15” x 15” x 24”
Weight 53lbs